:: Volume 19, Issue 3 (7-2021) ::
Int J Radiat Res 2021, 19(3): 695-702 Back to browse issues page
Analyzing the performance of the URT treatment planning system and URT-Linac 506C for flattening filter (FF) and flattening filter free (FFF) photon beams of Monte Carlo algorithm by use of AAPM TG 119 test cases
D. Jiang , D. Wang , J. Shen , J. Zhang , C. Chen , Z. Bao , H. Zhao , J. Chen , H. Liu
Department of Radiation and Medical Oncology, Hubei Key Laboratory of Tumor Biological Behaviors, Hubei Cancer Clinical Study Center, Zhongnan Hospital of Wuhan University, Wuhan, China, 430071 , hbzkznyy@163.com
Abstract:   (2091 Views)
Background: The objective in this study was to create AAPM TG 119 test plans for Intensity-modulated radiation therapy (IMRT), burst mode and volumetric modulated arc therapy (VMAT) in convolution study in order to investigate accuracy of the United Imaging Healthcare's URT treatment planning system (URT-TPS). Materials and Methods: The plans were delivered to the phantom using the United Imaging Healthcare's URT-Linac 506C. For there treatment mode as IMRT, Burst Mode and VMAT, with two kind of beams as flattening filter (FF) photon beam and flattening filter free (FFF) photon beam, calculated by uRT-TPS Monte Carlo algorithm, the overall accuracy was measured, and analyzed with five test geometries provided in TG 119. The point measurements were measured by a Farmer type ion chamber and fluence measurements were done with film respectively. Results: For the FF photon beams, the difference between measured point doses and planned doses of static multi-leaf collimator (MLC), dynamic MLC, Burst Mode and VMAT were within±3.92%,±3.26%, ±4.11%and±3.31% respectively. Gamma passing rates of Static IMRT, Dynamic IMRT Burst Mode and VMAT were >93.08%, 90.93%, 90.40% and> 92.00% respectively. For the FFF photon beams, the deviation between measured point doses and planed dose of static MLC, dynamic MLC, Burst Mode and VMAT were within 1.84%, 3.36%, 2.65%and 3.11% respectively. Gamma passing rates of Static IMRT, Dynamic IMRT Burst Mode and VMAT were>92.60%, 94.07%, 93.54% and 94.39% respectively and all confidence limits of the TG 119 report were matched. Conclusion: Based on this analysis which were performed in accordance with the TG 119 recommendations, it is evident that the URT treatment planning system and URT-Linac 506C have commissioned IMRT and VMAT techniques with adequate accuracy.
Keywords: TG 119, dosimetric validation, linac 506C, URT_TPS, flattening filter, flattening filter free
Full-Text [PDF 1912 kb]   (1208 Downloads)    
Type of Study: Original Research | Subject: Radiobiology
Send email to the article author

Add your comments about this article
Your username or Email:

CAPTCHA



XML     Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Jiang D, Wang D, Shen J, Zhang J, Chen C, Bao Z, et al . Analyzing the performance of the URT treatment planning system and URT-Linac 506C for flattening filter (FF) and flattening filter free (FFF) photon beams of Monte Carlo algorithm by use of AAPM TG 119 test cases. Int J Radiat Res 2021; 19 (3) :695-702
URL: http://ijrr.com/article-1-3793-en.html


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
Volume 19, Issue 3 (7-2021) Back to browse issues page
International Journal of Radiation Research
Persian site map - English site map - Created in 0.06 seconds with 50 queries by YEKTAWEB 4660