Usefulness of a head rotational correction device utilizing optical surface imaging in surface-guided radiation therapy
D. Kim , K.H Cho , J.H Jung , S. Kim
Department of Radiological Science, Gachon University Medical Campus, Incheon, Republic of Korea , ksc@gachon.ac.kr
Abstract:   (496 Views)
Background: Precise patient positioning is essential to effective head radiotherapy. However, helical tomotherapy lacks six-degree-of-freedom (6DoF) correction, limiting its ability to adjust for pitch and yaw errors. Optical surface imaging (OSI) enables non-invasive, real-time monitoring of patient position but exhibits limitations. Materials and Methods: A rotational correction device (RCD) was developed consisting of acrylic and steel adjustment components integrated into a carbon-fiber head-fixation baseplate. The phantom head is immobilized using a thermoplastic mask and positioned on the RCD. The setup accuracy and time were compared between the megavoltage computed tomography (MVCT)-based image-guided correction and OSI-assisted correction using an RCD. Ten trials were conducted for each system to measure the setup errors (translational and rotational magnitudes) and the total setup time. Treatment plans with a yaw error of 3° assessed the dosimetric impact on three spherical clinical target volumes (CTVs) and compared them with plans corrected using the RCD.  Results: The RCD system significantly reduced setup errors (0.24° ± 0.16° vs. 0.49° ± 0.14°; p=0.001) and setup time by approximately 40% (7 min 48 s vs. 19 min 37 s; p<0.001) compared with MVCT. Dosimetric analysis revealed improved D99% recovery in CTVs with RCD correction (e.g., CTV3: 24.74 Gy with RCD vs. 23.89 Gy without correction, against a reference of 24.74 Gy). Conclusion: The RCD could provide an effective and accurate solution for correcting pitch and yaw rotational errors during head radiotherapy. This approach enhanced setup precision and efficiency without requiring repeated imaging, thereby improving treatment accuracy in multitarget brain irradiation.
Keywords: Surface-guided radiation therapy, megavoltage computed tomography, tomotherapy, patient setup, rotational errors.
Full-Text [PDF 1052 kb]   (110 Downloads)    
Type of Study: Original Research | Subject: Radiation Biology
References
1. 1. Verellen D, Ridder MD, Linthout N, Tournel K, et al. (2007) Innovations in image-guided radiotherapy. Nat Rev Cancer, 7: 949-960. [DOI:10.1038/nrc2288]
2. Roper J, Chanyavanich V, Betzel G, Switchenko J, Dhabaan A (2014) TH-A-9A-11: Single-isocenter multiple-target SRS: Risk of compromised coverage. Medical Physics, 41: 545-545. [DOI:10.1118/1.4889581]
3. Guckenberger M, Roesch J, Baier K, Sweeney RA, Flentje M (2012) Dosimetric consequences of translational and rotational errors in frame-less image-guided radiosurgery. Radiat Oncol, 7: 63. [DOI:10.1186/1748-717X-7-63]
4. Gevaert T, Verellen D, Engels B, Depuydt T, et al. (2012) Clinical evaluation of a robotic 6-degree of freedom treatment couch for frameless radiosurgery. Int J Radiat Oncol Biol Phys, 83: 467-474. [DOI:10.1016/j.ijrobp.2011.05.048]
5. Hoogeman MS, Nuyttens JJ, Levendag PC, Heijmen BJM (2008) Time dependence of intrafraction patient motion assessed by repeat stereoscopic imaging. Int J Radiat Oncol Biol Phys, 70: 609-618. [DOI:10.1016/j.ijrobp.2007.08.066]
6. Boswell SA, Jeraj R, Ruchala KJ, Olivera GH, et al. (2005) A novel method to correct for pitch and yaw patient setup errors in helical tomotherapy: Correcting for rotational patient setup errors in tomotherapy. Med Phys, 32: 1630-1639. [DOI:10.1118/1.1914543]
7. Freislederer P, Kügele M, Öllers M, Swinnen A, et al. (2020) Recent advances in surface guided radiation therapy. Radiat Oncol, 15: 187. [DOI:10.1186/s13014-020-01629-w]
8. Hattel SH, Andersen PA, Wahlstedt IH, Damkjaer S, et al. (2019) Evaluation of setup and intrafraction motion for surface guided whole-breast cancer radiotherapy. J Appl Clin Med Phys, 20: 39-44. [DOI:10.1002/acm2.12599]
9. Haraldsson A, Ceberg S, Ceberg C, Bäck S, et al. (2020) Surface‐guided tomotherapy improves positioning and reduces treatment time: A retrospective analysis of 16 835 treatment fractions. J Appl Clin Med Phys, 21: 139-148. [DOI:10.1002/acm2.12936]
10. Wei W, Ioannides PJ, Sehgal V, Daroui P (2020) Quantifying the impact of optical surface guidance in the treatment of cancers of the head and neck. J Appl Clin Med Phys, 21: 73-82. [DOI:10.1002/acm2.12867]
11. Covington EL, Fiveash JB, Wu X, Brezovich I, et al. (2019) Optical surface guidance for submillimeter monitoring of patient position during frameless stereotactic radiotherapy. J Appl Clin Med Phys, 20: 91-98. [DOI:10.1002/acm2.12611]
12. Cravo Sá A, Fermento A, Neves D, Ferreira S, et al. (2018) Radiotherapy setup displacements in breast cancer patients: 3D surface imaging experience. Rep Pract Oncol Radiother, 23: 61-67. [DOI:10.1016/j.rpor.2017.12.007]
13. Hoisak JDP and Pawlicki T (2018) The Role of Optical Surface Imaging Systems in Radiation Therapy. Seminars in Radiation Oncology, 28: 185-193. [DOI:10.1016/j.semradonc.2018.02.003]
14. Swinnen ACC, Öllers MC, Loon Ong C, Verhaegen F (2020) The potential of an optical surface tracking system in non-coplanar single isocenter treatments of multiple brain metastases. J Appl Clin Med Phys, 21: 63-72. [DOI:10.1002/acm2.12866]
15. Lee SK, Huang S, Zhang L, Ballangrud AM, et al. (2021) Accuracy of surface‐guided patient setup for conventional radiotherapy of brain and nasopharynx cancer. J Applied Clin Med Phys, 22: 48-57. [DOI:10.1002/acm2.13241]
16. Usui K, Isobe A, Hara N, Muroi T, et al. (2019) Development of a rotational set‐up correction device for stereotactic head radiation therapy: A performance evaluation. J Appl Clin Med Phys, 20: 206-212. [DOI:10.1002/acm2.12616]
17. Beltran C, Pegram A, Merchant TE (2012) Dosimetric consequences of rotational errors in radiation therapy of pediatric brain tumor patients. Radiotherapy and Oncology, 102: 206-209. [DOI:10.1016/j.radonc.2011.06.013]
18. Dhabaan A, Schreibmann E, Siddiqi A, Elder E, Fox T, Ogunleye T, Esiashvili N, Curran W, Crocker I, Shu H-K (2012) Six degrees of freedom CBCT-based positioning for intracranial targets treated with frameless stereotactic radiosurgery. J Appl Clin Med Phys, 13: 215-225. [DOI:10.1120/jacmp.v13i6.3916]
19. Mancosu P, Reggiori G, Gaudino A, Lobefalo F, et al. (2015) Are pitch and roll compensations required in all pathologies? A data analysis of 2945 fractions. BJR, 88: 20150468. [DOI:10.1259/bjr.20150468]
20. Park S-Y, Choi N, Jang NY (2021) Frameless immobilization system with roll correction for stereotactic radiosurgery of intracranial brain metastases. Journal of Radiation Research rrab071. [DOI:10.1093/jrr/rrab071]
21. Ostyn M, Wang S, Kim Y-S, Kim S, Yeo W-H (2019) Radiotherapy-compatible robotic system for multi-landmark positioning in head and neck cancer treatments. Sci Rep, 9: 14358. [DOI:10.1038/s41598-019-50797-7]
22. Liu X, Belcher AH, Grelewicz Z, Wiersma RD (2015) Robotic real-time translational and rotational head motion correction during frameless stereotactic radiosurgery: Robotic real-time translational and rotational head motion correction during frameless stereotactic radiosurgery. Med Phys, 42: 2757-2763. [DOI:10.1118/1.4919279]
23. Roper J, Chanyavanich V, Betzel G, Switchenko J, Dhabaan A (2015) Single-isocenter multiple-target stereotactic radiosurgery: Risk of compromised coverage. Int J Radiat Oncol Biol Phys, 93: 540-546. [DOI:10.1016/j.ijrobp.2015.07.2262]
24. Tryggestad E, Christian M, Ford E, Kut C, et al. (2011) Inter- and intrafraction patient positioning uncertainties for intracranial radiotherapy: A study of four frameless, thermoplastic mask-based immobilization strategies using daily cone-beam CT. Int J Radiat Oncol Biol Phys, 80: 281-290. [DOI:10.1016/j.ijrobp.2010.06.022]
25. Tarnavski N, Engelholm SA, Rosenschold PMAf (2016) Fast intra-fractional image-guidance with 6D positioning correction reduces delivery uncertainty for stereotactic radiosurgery and radiotherapy. J Radiosurg SBRT, 4(1): 15-20.
26. Mackie TR, Holmes T, Swerdloff S, Reckwerdt P, et al. (1993) Tomotherapy: a new concept for the delivery of dynamic conformal radiotherapy. Med Phys, 20: 1709-1719. [DOI:10.1118/1.596958]
27. Park H-K, Cho J-H, Kim S (2019) Dosimetric Effects of Rotational Setup Error in Volumetric Modulated Arc Radiotherapy on Brain Tumor Patients. Iran J Med Phys, 16: 171-178.
28. Prentou G, Pappas EP, Logothetis A, Koutsouveli E, et al. (2020) Dosimetric impact of rotational errors on the quality of VMAT‐SRS for multiple brain metastases: Comparison between single‐ and two‐isocenter treatment planning techniques. J Appl Clin Med Phys, 21: 32. [DOI:10.1002/acm2.12815]
29. Zhao B, Maquilan G, Jiang S, Schwartz DL (2018) Minimal mask immobilization with optical surface guidance for head and neck radiotherapy. J Appl Clin Med Phys, 19: 17-24. [DOI:10.1002/acm2.12211]
30. Zhang L, Vijayan S, Huang S, et al. (2021) Commissioning of optical surface imaging systems for cranial frameless stereotactic radiosurgery. J Applied Clin Med Phys, 22: 182-190. [DOI:10.1002/acm2.13240]
31. Wiencierz M, Kruppa K, Lüdemann L (2016) Clinical validation of two surface imaging systems for patient positioning in percutaneous radiotherapy. ArXiv:1602.03749v1 [physics.med-ph]
32. Kang HJ, Grelewicz Z, Wiersma RD (2012) Development of an automated region of interest selection method for 3D surface monitoring of head motion. Med Phys, 39: 3270-3282. [DOI:10.1118/1.4711805]
33. Xu G, Xiong Z, Wang H, Jiang H, et al. (2018) A feasibility study of dynamic verification for tumor target delineation and dose delivery using a six degrees of freedom motion phantom. Int J Radiat Res, 16(4): 411-420.
Send email to the article author

Add your comments about this article
Your username or Email:

CAPTCHA



XML     Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Kim D, Cho K, Jung J, Kim S. Usefulness of a head rotational correction device utilizing optical surface imaging in surface-guided radiation therapy. Int J Radiat Res 2026; 24 (4) :1081-1086
URL: http://ijrr.com/article-1-7386-en.html


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
Volume 24, Issue 4 (10-2026) Back to browse issues page
International Journal of Radiation Research
Persian site map - English site map - Created in 0.14 seconds with 48 queries by YEKTAWEB 4741